Recall Z-0462-2011 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

Signa Excite" 1.5 T Magnetic Resonance Systems, manufactured by GE Healthcare, P.O. Box 44 Milwaukee, WI 53201.

DateNov 24, 2010 (initiated May 11, 2009)
StatusTerminated
ReasonSigna MR Excite" 1.5T system containing 11.1_M4_0818.a software, associated with 2D Fast Spin Echo based pulse sequences (2D FSE-XL, 2D FRFSE-XL, 2D FSE-IR, 2D T2FLAIR and 2D T1FLAIR) may experience an image flip along the phase encode direction and incorrect annotation, which may result in incorrect localization of anatomy in oblique axial images with the pulse sequences.
Root causeSoftware design
ActionGE Healthcare issued an "Urgent Medical Device Correction" letter dated April 21, 2009. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. GE Field Engineers will contact customers to schedule a field modification of the affected products to address this issue. Customers can contact GE Healthcare at 262 521-6100 about this issue.
Quantity1464
DistributionWorldwide Distribution: Throughout USA except states of AK, VT and WY, and the countries of: Argentina, Australia, Bulgaria, Brazil, China, Canada, Switzerland, Chile, Colombia, Czech Republic, Ecuador, Egypt, Spain, Finland, France, Great Britain, Guatemala, Honduras, Croatia, Hungary, Indonesia, I
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013636SIGNA INFINITY MR SYSTEM WITH EXCITE TECHNOLOGY
GE Medical Systems, Jan 24, 2002