ClearanceBrief
Recall Z-1368-2011 ยท IZI System, X-Ray, Angiographic

Philips Medical Systems North America Co. Phillips

Allura XPER FD 10 and Field Extension for XPER Cardiovascular System and Allura XPER FD 20 and Allura CV20. Angiographic X-ray systems with a solid state X-ray imaging device for cardiovascular and vascular diagnostic and interventional procedures.

DateMar 22, 2011 (initiated Nov 4, 2010)
StatusTerminated
ReasonAdjustment of the micro-switches in the footswitch used on the Philips cardiovascular X-ray systems was not implemented. If it is programmed for exposure, it is possible that after pressing the pedal on the footswitch it does not release itself. The patient may be exposed to X-ray longer than intended.
Root causePending
ActionOn 11/5/10, Philips began sending the Electronic Product Radiation Warning notification letter, dated November 4, 2010, to their consignees. The consignees were informed that the middle pedal of the foot switch of the Allura FD10, FD20, CV20 systems, and the Allura Field Extension series does not not always become inactive (sticks) when pedal is released. The function of the middle pedal on the foot switch is freely programmable by the customer. If it is programmed for exposure and if the pedal does not release by itself, the patient will be exposed to X-ray longer than intended. The co
Quantity4 units
DistributionWorldwide Distribution -- US, including states of CA, IL, MA, NC, and OC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033737ALLURA XPER FD20
Philips Medical Systems North America, Inc., Dec 9, 2003
K041949ALLURA XPER FD10
Philips Medical Systems North America Co., Jul 30, 2004