Recall Z-1475-2011 ยท MQP Spinal Vertebral Body Replacement Device

Integra LifeSciences Corp.

Vu a-POD Intervertebral Body Fusion Device. Packaged as surgical kits containing all implants and instruments required for a procedure. Also packaged as individual instruments and implants in poly bags distributed as replenishment for surgical kits. Theken Spine, LLC, Akron, OH 44306. Indicated for use as an adjunct for fusion in patients with degenerative disc disease (DDD) at one or two cont

DateMar 2, 2011 (initiated Oct 19, 2010)
StatusTerminated
ReasonThe IFU provided with the device contains incorrect information in the Indications for Use Section, stating the following: The Theken Spine Vu a"POD Intervertebral Body Fusion Device is intended for use as a stand alone implant when used with the SpinPlate provided in the set.
Root causeEmployee error
ActionOn 10/19/2010 the firm sent "URGENT : FIELD CORRECTION ACTION Letters to their customers. The letters identified the affected product and explained the reason for the recall. The letter also stated that the issue is not believed to have a health impact on patients. Physicians are asked to continue to exercise their clinical judgment with any periodic patient follow up. Attached to the letters were IFUs containing the correct information. Customers are to complete and return the enclosed Recall Acknowledgement Form per the instructions provided. Questions can be directed to Suzanne B. Woj
Quantity2,163 pieces
DistributionNationwide Distribution -- AL, AZ, CA, CT, ID, KS, MO, NH, NV, NY, OH, OK, TX & UT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080822VU APOD INTERVERTABRAL BODY FUSION DEVICE
Theken Spine, LLC, Jul 2, 2008