Recall Z-0837-2011 ยท FRN Pump, Infusion

B Braun Medical, Inc

Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump IV Sets

DateJan 10, 2011 (initiated Aug 24, 2010)
StatusTerminated
ReasonPossible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive, ceasing normal operation.
Root causeComponent design/selection
ActionB. Braun Medical Inc. notified consignee by letter on August 24, 2010.
Quantity2,107 Pumps (582 Pumps to Winchester)
DistributionVirginia and New York
FDA recordOpen on fda.gov