Recall Z-0600-2011 ยท NZW Heparin, Vascular Access Flush

Medefil Incorporated

Medefil Heparin I.V. Flush Syringe 10 units/mL; a polypropylene luer locking syringe containing Heparin Lock Flush Solution, USP, a sterile non-pyrogenic, isotonic injection; each mL in a Heparin I.V. Flush Syringe contains 10 units of heparin sodium derived from porcine intestinal mucosa and 9 mg of sodium chloride in water for injection; Medefil, Inc. Glendale Heights, IL 60139. The pre-filled

DateDec 10, 2010 (initiated Nov 1, 2010)
StatusTerminated
ReasonThe heparin lock flush solution was manufactured with contaminated Heparin Sodium USP Active Pharmaceutical Ingredient (API).
Root causeMaterial/Component Contamination
ActionMedefil faxed/emailed the direct accounts on 11/1/10, and sent the accounts a hard copy of the Urgent Product Recall letter, dated 11/1/2010, on 11/2/10. The accounts were informed of Medefil's recall of 18 lots of Heparin I.V. Flush Syringes produced from the active pharmaceutical ingredient (API) Heparin Sodium, USP, lot 1035-0780, that has been recalled by the manufacturer, Scientific Protein Laboratories, due to the presence of a contaminant, Oversulfated Chondroitin Sulfate. The accounts were instructed to stop use of the 18 lots of Heparin I.V. Flush Syringes immediately, place them un
Quantity583,500 syringes
DistributionNationwide Distribution --- including states of Arizona, California, Florida, Illinois, Ohio, Louisiana, Massachusetts, Michigan, Minnesota, New York, Texas, and Virginia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092491MEDEFIL'S HEPARIN I.V. FLUSH SYRINGE 1 UNIT/ML; 10 UNITS/ML; AND 100 UNITS/ML IN VARIOUS FILL SIZES
Medefil, Inc., Feb 3, 2010