Recall Z-0544-2011 ยท IZI System, X-Ray, Angiographic

Philips Medical Systems North America Co. Phillips

MultiDiagnost Eleva with FD. The Philips MultiDiagnost Eleva intended use is for the following applications: As a diagnostic-imaging device multi-functional/ universal system, general R/F, Fluoroscopy, Radiography and Angiography can be performed along with pediatric examinations and some more specialized interventional applications.

DateDec 7, 2010 (initiated May 26, 2009)
StatusTerminated
ReasonThe Printed Circuit Board (PCB) in the Main-cabinet. Further investigation revealed that a manufacturing error in the PCB is responsible for the reliability issue. As a result, the complete system may shut down and not re-start.
Root causeOther
ActionPhilips began mailing out an URGENT-Device Correction letter on May 26, 2009. Philips sent out a second URGENT-Device Correction letter dated June 4, 2009, to all consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were advised to contact their local Philips representative immediately if their system completely shut down and would not restart. A mandatory Field Change Order 72200144 was issued for replacement of the Printed Circuit Board (PCB) in the Allura XPER FD system. A mandatory Field Change Order FCO 70800091 was issu
Quantity5 units
DistributionNationwide Distribution including the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, OR, PA, RI, TN, TX, VA, and WI.
FDA recordOpen on fda.gov