Recall Z-0411-2011 ยท OLO Orthopedic Stereotaxic Instrument

Smith & Nephew Inc

smith & nephew TRIGEN SURESHOT Distal Targeting System, Trauma Interface, V2.0.2, Trauma Launcher V1.01.1 Product is intended to be an intraoperative image-guided localization system. It is a computer-assisted orthopedic surgery tool to aid the surgeon with drill positioning for screws during intramedullary nail implantation.

DateNov 18, 2010 (initiated Oct 15, 2010)
StatusTerminated
ReasonTargeting software has been found to provide erroneous targeting when used with the 13 mm TAN/FAN Left nails.
Root causeOther
ActionSmith & Nephew Inc. sent an URGENT MEDICAL DEVICE CORRECTION letter by FedEx on October 15, 2010, to all customers. The letter was dated October 4, 2010, and identified the product, the problem, and the action to be taken by the customers. Customers were instructed to perform a software upgrade and complete the form provided and return by fax to (901) 566-7835. For questions regarding this recall call (901) 399-5441.
Quantity445 units
DistributionWorldwide Distribution - USA including AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NM, OH, PA, TN, TX, AND WA and the countries of Australia, Belgium, Canada, China, Denmark, Finland, France, Germany, Italy, Japan, Netherlands, Norway, South Africa, Spain, Sweden,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100107SMITH & NEPHEW SURESHOT DISTAL TARGETING SYSTEM V2.0
Smith & Nephew, Inc., Feb 23, 2010