Recall Z-0414-2011 ยท DXJ Display, Cathode-Ray Tube, Medical
Philips Healthcare Inc.
Philips IntelliVue Clinical Information Portfolio (ICIP), Clinical Management System Software Version D.00 through D.03.
| Date | Nov 18, 2010 (initiated Oct 19, 2010) |
| Status | Terminated |
| Reason | Software: Scheduling function of the Philips Intellivue Clinical Information Portfolio Releases D.00 through D.03. Edits to a scheduled even frequency order may result in pending interventions not being generated after the first 24 hours. |
| Root cause | Software design |
| Action | Philips Healthcare issued an Urgent Medical Device Correction letter dated October 15, 2010 to customers. Customers are asked to follow the guidelines in the "Action to be taken by Customer/User" section of the Urgent Medical Device Correction Notice until they receive their software correction. The correction will be in the form of a software upgrade and will be performed by Philips service personnel. Contact Philips Customer Care Center with questions or concerns about this correction at 1-800-722-9377 |
| Quantity | 200 units |
| Distribution | Worldwide Distribution: USA, including states of CA, DC, FL, IL, IN, KY, MA, MD, ME, NJ, NY, OK, RI, TX, VA, VT, and WV; and the countries of: AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CHINA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, HONG KONG, IRELAND, ITALY, KENYA, LATVIA, MALAYSIA, NETHERLAND |
| FDA record | Open on fda.gov |