Recall Z-0417-2011 ยท INI Vehicle, Motorized 3-Wheeled

Electric Mobility Corp

Rascal UltraLite 370B manufactured by Giant Manufacturing, Taiwan.

DateNov 18, 2010 (initiated Sep 27, 2010)
StatusTerminated
ReasonScooter seat bases were found to have insufficient welding which may lead to the seat base becoming loose and causing the seat to wobble. This may result in complete failure of the seat base, causing the user to fall off the vehicle.
Root causeDevice Design
ActionThe Rascal Company (Electric Mobility Corporation) sent Urgent Medical Device Recall Notice Scooter Seat Base letters dated October 7, 2010 and November 11, 2010 letter to their affected retail customers and affected dealers. The letter instructs the reader that the seat base may become loose, causing the seat to wobble and if not replaced the wobble could lead to a complete failure of the seat base and cause an individual to fall off their vehicle and become injured. The letters also instruct the reader to replace seat bases immediately. Respone forms are attached. Seats with new seat bases
Quantity315 units
DistributionWorldwide Distribution: Throughout the US and Canada