Recall Z-0490-2011 ยท JSS Kit, Identification, Enterobacteriaceae

AdvanDx, Inc.

AdvanDx E. coli/P. aeruginosa PNA FISH Culture Identification Kit for the diagnosis of E. coli and/or P. aeruginosa bacteremia Catalog Number: KT007

DateNov 29, 2010 (initiated Oct 6, 2010)
StatusTerminated
ReasonStrains of Acinetobacter radioresistens cross-react with both E. coli/P. aeruginosa PNA FISH and EK/P. aeruginosa PNA FISH to produce a false positive red signal.
Root causeDevice Design
ActionAdvanDx issued a customer letter dated October 6, 2010 with a PDF file of a new package insert and an acknowledgement form to fill out and return to AdvanDx. Customers were instructed to replace the package insert in any existing inventory with the enclosed package insert. Customers can contact AdvanDx technical support about this action at 1-866-376-0009.
Quantity260 units
DistributionWorldwide Distribuition: Throughtout the US, and to the countries of Denmark, Italy, Sweden, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092236E.COLI/P. AERUGINOSA PNA FISH
Advandx, Inc., Dec 16, 2009