Recall Z-1592-2011 ยท LLZ System, Image Processing, Radiological

Straumann Manufacturing Inc.

Straumann coDiagnostiX, dental implant planning system. Software Version 7.0 including updates 7.0.2 and 7.0.2.10

DateMar 7, 2011 (initiated Sep 30, 2010)
StatusTerminated
ReasonSoftware: Surgical Error may occur when the implant length is changed after positioning the surgical sleeve.
Root causeSoftware design
ActionStraumann notified accounts of this correction by letter dated September 30, 2010 and delivered via Federal Express Delivery Service. The upgrade is available on-line by logging into the coDiagnostiX system and enabling the automatic upgrade to occur.
Quantity20 systems
DistributionNationwide