Recall Z-0816-2011 ยท FMI Needle, Hypodermic, Single Lumen

Becton Dickinson & Company

BD Vacutainer Push Button collection Set; Non-Pyrogenic, Sterile, Rx only; 23 G x 3/4 x 12" (0.6 x 19 mm x 305 mm) Becton Dickinson & Co., Franklin Lakes, NJ, USA Intended use: Blood collection

DateDec 27, 2010 (initiated Oct 5, 2010)
StatusTerminated
ReasonThere is the possibility for certain lots of the BD Vacutainer Push Button Blood Collection sets to have separation of the cannula from the hub.
Root causeProcess control
ActionBD Diagnostics sent "RECALL NOTIFICATION" separate letters October 1 and 5, 2010 and follow up letter dated October 8, 2010, as well as a letter dated December 1, 2010 to all distributors and customers commencing October 5, 2010 by federal express. The letters describe the product, the problem and the action to be taken by the customers. The customers were instructed to immediately examine their inventory for specific catalog and lot numbers, and return product from the lots following the instructions on the enclosed packing slip (BD will send replacement upon receipt of returned product). If
Quantity4,223,200
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011984BD VACUTAINER PUSH BUTTON BLOOD COLECTION SET
Becton, Dickinson & CO, Aug 29, 2001