Recall Z-0409-2011 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
GE Healthcare, LLC
Innova 3100/3100IQ GE Innova 3100 / 3100 IQ, Cardiovascular Imaging System, GE Healthcare, Waukesha, WI 53188 (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector) The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and option
| Date | Nov 17, 2010 (initiated Aug 5, 2009) |
| Status | Terminated |
| Reason | GE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy availability for approximately 5 minutes should there be a main power failure. Once power is restored, the equipment power will be returned to the main hospital power. UPS fluoro ba |
| Root cause | Nonconforming Material/Component |
| Action | GE Healthcare sent an "Urgent Medical Device Correction" letter on August 5, 2009 to all Hospital Administrators / Risk Mangers, Mangers of Radiology / Cardiology and Radiologists / Cardiologists. The letter was dated August 4, 2009, and described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. For questions regarding this recall call (262) 548-2731. |
| Quantity | 207 |
| Distribution | Worldwide Distribution - USA including AL, AZ, AR, CA, CO, CT, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK ,OR, PA, SC, TN, TX, VT, VA, WA, WV, and WI and the countries of ANGOLA, BELARUS, BULGARIA, CANADA, EGYPT, FRANCE, GERMANY, HUNGARY, I |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K052157 | INNOVA CT OPTION FOR DIGITAL FLUOROSCOPIC IMAGING SYSTEMS INNOVA MODELS 4100, 4100IQ, 3100, 3100IQ & 2100IQ Ge Medical Systems, LLC, Aug 25, 2005 |
| K052412 | INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQ GE Healthcare, Sep 16, 2005 |
| K061163 | INNOVA 2121IQ, INNOVA 3131IQ, INNOVA 4100, INNOVA 4100IQ, INNOVA 3100, INNOVA 3100IQ, INNOVA 2100IQ WITH INNOVA IVUS OPT Ge Medical Systems, LLC, Jun 7, 2006 |