Recall Z-0361-2011 ยท GDW Staple, Implantable

Ethicon Endo-Surgery Inc

ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Model # ATS45. Manfuactured by Ethicon Endo-Survery, LLC; 475 Calle C; Guaynabo PR 00969.

DateNov 15, 2010 (initiated Jan 13, 2009)
StatusTerminated
ReasonThe mechanism that connects the articulation joint to the device shaft may cause the jaws of the cutters to remain closed and clamped down on tissue after the device is fired. If the jaws of the device remain clamped down on the tissue, there is a risk the surgery will need to be modified to remove the device. This may lead to an increase in procedure time, tissue manipulation, and a change in t
Root causeOther
ActionEthicon Endo-Surgery issued "Urgent Device Recall Event 1952" notification letters dated January 21, 2009 to their customers. They instructed their custormers to immediately discontinue use of the product. They are to complete and fax back the Business Reply Form. Affected product is to be returned with a copy of the completed Business Reply Form to Stericycle. Replacement product should be received within 30-60 days after return. Customers can contact Stericycle concerning this recall at 1-866-861-6921, reference Event 1952.
Quantity4,113 UNITS
DistributionWorldwide Distribution: USA only, in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI. IA. ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020779PROXIMATE AND ENDOPATH LINEAR CUTTERS AND STAPLERS
Ethicon Endo-Surgery, Inc., Jun 5, 2002
K070887ENDOPATH LINEAR CUTTERS AND STAPLERS AND ECHELON60 ENDOSCOPIC LINEAR CUTTERS AND STAPLERS (AND RELOADS)
Ethicon Endo-Surgery, LLC, May 25, 2007