Recall Z-0927-2011 ยท LXH Orthopedic Manual Surgical Instrument

Stryker Howmedica Osteonics Corp.

Scorpio Adjustable Femoral A/P Sizing Guide Assembly; non-sterile instruments; Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430; A subsidiary of Stryker Corporation The A/P sizer is designed to set the desired external rotation and to provide adjustment of the anterior/posterior position when needed.

DateJan 19, 2011 (initiated Sep 27, 2010)
StatusTerminated
ReasonStryker Orthopaedics has become aware that there is a potential for the shoulder screw on the Scorpio Sizing Guide (the "Sizing Guide") to disassociate from the main body of the instrument.
Root causeDevice Design
ActionStryker Orthopaedics sent an 'Urgent Update Product Recall' letter dated December 30, 2010, along with Product Acknowledgement forms to all customers (US and international). The letter described the product, problem and action to be taken by the customers. The customers were asked to complete and return the PRODUCT RECALL ACKNOWLEDGMENT FORM within five (5) days via fax at 201-831-6069; contact the hospitals in their territory that have the affected product to arrange return of the product; and return the affected product to the attention of Regulatory Compliance, Stryker Orthopaedics, 325 C
Quantity237 units
DistributionWorldwide distribution: USA and countries including: Australia, Brazil, China, Columbia, France, Greece, India, Italy, Korea, Poland, Singapore, South Africa, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.