ClearanceBrief
Recall Z-0342-2011 ยท CAC Apparatus, Autotransfusion

Fresenius Kabi, LLC

The box label states: "ATF 120 Fast Start Kit 9108501...STERILE...FRESENIUS KABI...Fresenius Kabi AG 61345 Bad Hamburg Germany...Federal (USA) Law restricts this device to sale by or on order of a physician". The ATF120 Fast Start Kit contains one each AT1 Autotransfusion Set, ATS Suction Line, and the ATR120 Autotransfusion Reservoir. ATS is a suction line to aspirate and add anticoagulate to s

DateNov 10, 2010 (initiated Mar 29, 2010)
StatusTerminated
ReasonUsing AT1 Autotransfusion Set from specific batches in combination with the auto transfusion device CATS Continuous Autotransfusion has led to insufficient suction of shed blood from the collection reservoir. The results in the failure message "alarm blood flow" and the CATS device stops in a secure mode.
Root causeOther
ActionFresenius Kabi sent out a Technical Notification letter dated March 29, 2010, to one distributor located in Ann Arbor, Michigan. The distributor sent the technical notification with a cover letter to each of their customer's that had ordered AT1's, ATF120's and ATF40's over the last year. The letter identified the product, the problem, and the action to be taken. Customers were notified that the autotransfusion prime and / or washing procedure can be continued after alarm blood flow by following Alternative procedures. The Technical Notification informs the distributor about the proble
Quantity25, 740 units distributed in US.
DistributionWorldwide Distribution - USA and the countries of Australia,Austria, Brazil, Canada, China, Denmark, France, Germany, Hong Kong, India, Italy, Japan, Malaysia, Netherlands, Phillipines, Portugal, Switzerland, Thailand, UK, and Vietnam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K960006C.A.T.S. CONTINUOUS AUTOTRANSFUSION SYSTEM
Fresenius AG, Jul 31, 1996