Recall Z-0709-2011 ยท JEY Plate, Bone

Stryker Craniomaxillofacial Division

2.8 mm Inion CPS, 2.8 x 16 mm screw, REF SCR-1300, sterile, Manufacturer, INION Oy, Finland.

DateDec 16, 2010 (initiated Apr 14, 2010)
StatusTerminated
ReasonInadequate Instructions for Use. The stated strength retention is ambiguous and the maximum immersion time of the plates/meshed is not defined.
Root causeOther
ActionStryker issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 14, 2010. The notice informed users of the change in the instructions for use (IFU), and included a copy of the revised IFU. The consignees were informed that the affected product does not need to be returned. Stryker can be contacted about this action at 877-534-2464 (option "Regulatory Affairs", reference Product Field Action #2009-445)
DistributionNationwide Distribution: Throughout the US
FDA recordOpen on fda.gov