Recall Z-0203-2011 ยท LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation

Biomerieux Inc

VITEK 2 Gram Negative Susceptibility Card, AST-N093, REF 22 246, For In Vitro Diagnostic Use, 20 cards/box. The firm names on the label are bioMerieux, Inc., Durham, NC, and bioMerieux sa, 69280 Marcy-l'Etoile France.

DateNov 11, 2010 (initiated Sep 16, 2010)
StatusTerminated
ReasonAMENDED REASON FOR RECALL: The piperacillin/tazobactam drug on the antimicrobial susceptibility cards is providing false susceptible and false resistant results for E. coli, as well as an additional 7 organisms; and the same drug is providing false resistant results for K. pneumoniae.
Root causeNonconforming Material/Component
ActionbioMerieux issued a "Product Correction Notice" letter dated September 13, 2010, to their entire customer base (regardless of whether they ever ordered the recalled cards). The letters explained the reason for the recall and instructed the account to confirm all results for E. coli using a second method, and for K. pneumoniae, to confirm with a second method only for resistant results. These alternate test confirmations are to continue until further notice. The customers were to complete and return the enclosed Acknowledgement Form via fax at (314) 731-8689 or (919) 620-6350 which indicate
Quantity23,060 boxes as of 3/8/11
DistributionWorldwide Distribution: USA (all states), and the countries of: SUBSIDIARIES - Argentina, Canada, Switzerland, Chile, Colombia, Germany, Spain, Great Britain, Hong Kong, Japan, Korea, Mexico, Poland, Brazil, Sweden, Singapore, and Thailand. DISTRIBUTORS - Dutch Antilles, Australia, Bermuda, Urugu
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080107VITEK 2 GRAM NEGATIVE PIPERACILLIN/TAZOBACTAM
bioMerieux, Inc., Feb 28, 2008