Recall Z-0940-2011 ยท DXC Clamp, Vascular

Maquet Cardiovascular, LLC

Maquet Heartstring II Proximal Seal System (HS-1045) manufactured by Maquet Cardiovascular, Wayne, NJ Intended use: medical device for use by cardiac surgeons during CABG procedures to maintain hemostasis and to facilitate the completion of a proximal anastomosis without application of an aortic clamp.

DateFeb 1, 2011 (initiated May 5, 2010)
StatusTerminated
ReasonDeployment tube may detach during use due to insufficient adhesive, which may result in product failure to operate.
Root causeProcess design
ActionThe firm, MAQUET Getinge Group, sent an "URGENT DEVICE REMOVAL IMMEDIATE ACTION REQUIRED" letter dated May 5, 2010, to all customers via Federal Express. The letter described the product, problem and actions to be taken by the customer. The customers were instructed to immediately examine their stocks to determine if they have any of the affected product, if found, discontinue dispensing (distributing) the lot and complete the Field Action Response Form; immediately contact accounts if they acted as a distributor, and acknowledge receipt of this notification whether or not they have any affe
Quantity1630
DistributionWorldwide distribution: USA and countries including: Japan, Mexico, Germany, Switzerland, Israel
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022880GUIDANT PROXIMAL SEAL SYSTEM, MODELS PSS-4500 & PS-4500
Guidant Corp., Sep 13, 2002