ClearanceBrief
Recall Z-0924-2011 ยท JEA Table, Surgical With Orthopedic Accessories, Ac-Powered

Mizuho Orthopedic Systems Inc

Mizuho ProFx 6850 Operating Room Table, with orthopedic accessories, AC Powered, manufactured by Mizuho OSI, Union City, CA.

DateJan 20, 2011 (initiated Mar 30, 2010)
StatusTerminated
ReasonDevice component spar gear box locks up and may result in failure of push pull tube, which may result in release of breaking and impact.
Root causeDevice Design
ActionAll customers will be contacted to arrange for removal and shipping arrangements. Verification will be done by the Repair and Verification instructions for the leg spars, via registered letters. *******UPDATED INFO---The firm sent another notification letter---dated 12/22/10---which states in part: "****The upgrade involves the correction of your 6850 PROfx spars (6850-2100) at your facility by a Qualified Technician. The spars shall be upgraded to remove the 6850-2123, Drive Bracket, Spar Push Rod and replace it with the 6850-2170, Drive Bracket, Spar Push Rod. Upon completion of the upgrad
Quantity17 units
DistributionWorldwide Distribution -- USA & Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.