ClearanceBrief
Recall Z-0849-2011 ยท KQM Camera, Surgical And Accessories

Linvatec Corp. dba ConMed Linvatec

Video Cart Multi-Specialty product labeling is as follows: REF, VP6502, QTY1, Video Cart Multi-Specialty, 120V, Rx ONLY, These Video Carts are intended for storage of Endoscopic equipment that is used in single or multiple surgical specialties MANUFACTURED FOR: Linvatec, LINVATEC CORPORATION, LARGO, FL 33773 USA.

DateJan 4, 2011 (initiated Feb 23, 2010)
StatusTerminated
ReasonConMed Linvatec, Endoscopy Division, dba. ConMed Linvatec, Largo, Florida is recalling five products: (1.) Video Cart Premium 120 VAC (VP6501), (2.) Video Cart Multi-Specialty 120 VAC (VP6502), (3.) Video Cart Articulating Arm 120V (VP6504), (4.) Video Cart (VP8500), and (5.) Power Cord Assembly Self Retracting (VP6547). These products are being recalled due to there is a possibility that the Vid
Root causeComponent design/selection
ActionConMed Linvatec sent URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letters to their customers by FedEx. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to review their plugs for damage and determine if they were affected. Distributors were asked to contact affected customers and schedule delivery of their replacement plugs within 30 days. If the account no longer has the product in their possession, or if they have transferred the product, they were asked to respond as such via a Reply form. Customers were asked to document
Quantity173 carts
DistributionWorldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, AND WY and the countries of Canada, Chile, Colombia, Denmark, Korea, Mexico, Singapore, So
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.