Recall Z-0611-2011 ยท IKZ Bed, Patient Rotation, Powered

KCI USA, Inc.

RotoProne Therapy System, Model 209500, Device List Number E232632 Product is designed to allow patients to be mechanically rotated to the prone position and provides kinetic therapy to patients. It is intended for the treatment and prevention of pulmonary complications associated with immobility.

DateDec 13, 2010 (initiated Sep 9, 2010)
StatusTerminated
ReasonProduct complaints indicating that while attempting to resolve an alarm or error condition, operator was unable to return product to the prone position for patient.
Root causeOther
ActionKCI USA, Inc. sent an URGENT - Medical Device Safety Alert letter, dated September 8, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the cusotmer. Customers were instructed to have a practiced back-up plan, such as a plan for the manual proning of patients in the event that clinical circumstances dictate the need for more immediate prone therapy. For questions, customers in the US should call KCI's Technical Support Center at 1-800-275-4524, option 5. For customers outside the US they should contact their local sales represe
Quantity240 units
DistributionWorldwide Distribution - USA Nationwide, United Kingdom, Germany, and Saudi Arabia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.