Recall Z-0808-2011 ยท FTL Mesh, Surgical, Polymeric

American Medical Systems, Inc.

Model # 720088-01: AMS AdVance Male Sling System, US; Model # 720088-02: AMS AdVance Male Sling System, INTL; Model # 72404224: Male TO Sling System, INTL. Sterilization Method ETO, American Medical systems, Inc., 10700 Bren Road West, Minnetonka, MN 55343 USA

DateDec 27, 2010 (initiated Sep 7, 2010)
StatusTerminated
ReasonAmerican Medical Systems (AMS) is recalling all lots of the AdVance male Sling System with an expiration date between July 2012 and August 2013. Internal testing has indicated the potential for degradation of the bioabsorbable tensioning sutures when the product is stored for longer than 15 months after the date of manufacture.
Root causePackage design/selection
ActionConsignees were sent an AMS "Urgent Product Recall" letter dated September 7, 2010. The letter described the problem and the product involved. They requested consignees to return any existing stock product and to return the Product Recall Acknowledgement form enclosed. Customers may request a credit or a full refund or also, arrange for replenishment of AdVance product. Questions regarding the recall should be directed towards AMS Customer Service at 1-800-328-3881.
Quantity7,069
DistributionWorldwide Distribution -- All states in the Continental USA including DC and Puerto Rico and countries of ARGENTINA, AUSTRIA, AUSTRALIA, BELGIUM, BRAZIL, CANADA, SWITZERLAND, CHILE, CHINA, COLOMBIA, COSTA RICA, GERMANY, ESTONIA, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, GREECE, HONG KONG, CROATIA, IRE
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053371AMS MALE TRANSOBTURATOR SLING SYSTEM
American Medical Systems, Inc., Feb 3, 2006