Recall Z-0368-2011 ยท JAK System, X-Ray, Tomography, Computed

GE Healthcare, LLC

GE LightSpeed VCT Scanner System, models 5212920-300, 5212920-310, 5311595-10. Composed of a gantry, patient table, operator console, computer, and PDU and includes image acquisition hardware, image acquisition and reconstruction software, associated accessories and connections/interfaces to accessories. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188 USA. Is intended to produce cross

DateNov 15, 2010 (initiated Sep 3, 2010)
StatusTerminated
ReasonGE has become aware of a potential set of circumstances that could cause x-ray continuation during an unexpected table stop on certain LightSpeed VCT and BrightSpeed systems.
Root causeSoftware design
ActionAn Urgent Medical Device Correction letter dated September 2, 2010 was sent to consignees beginning September 3, 2010. This letter addressed the LightSpeed VCT systems. Another Urgent Medical Device Correction letter, dated September 24, 2010 was sent September 24, 2010 for the BrightSpeed Systems. The letters described the issue and provided safety instructions for the health care provider to be aware if the table should stop unexpectedly and the x-ray remains on to manually stop the scan until the devices software is upgraded. If you have any questions, please contact the US (800) 437-1
Quantity68
DistributionWorldwide distribution: USA including states of: MI, NM, OH, TX, and Puerto Rico and countries including: Algeria, Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Denmark, Dominican Republic, Ecuador
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052855STARSPEED SERIES CT SCANNER SYSTEMS
GE Healthcare, Oct 28, 2005
K082761GE LIGHTSPEED 7.2 CT SCANNER SYSTEM
Ge Medical Systems, LLC, Oct 8, 2008
K082816GE BRIGHTSPEED DELIGHT CT SCANNER SYSTEM
Ge Medical Systems, LLC, Oct 8, 2008