ClearanceBrief
Recall Z-0092-2011 ยท MND Ligator, Esophageal

Cook Endoscopy

Six Shooter Saeed Multi-Band Ligator, Rx Only, Wilson Cook Medical 4900 Bethania Station Road, Winston-Salem, NC 27105 This device is used to endoscopically ligate esophageal varices at or above the gastroesophageal junction or to ligate internal hemorrhoids.

DateOct 19, 2010 (initiated Sep 3, 2010)
StatusTerminated
ReasonA section of the ligator barrel may separate from the barrel's friction fit adapter that attaches to the patent end of the endoscope. This could result in a section of the ligator barrel detaching inside the patient's gastrointestinal tract.
Root causeOther
ActionCOOK Endoscopy sent an URGENT: PRODUCT RECALL dated September 3, 2010, via USPS Certified Mail and by email to all consignees. The letter identified the products, the problem, and the action to be taken by the consignees. All consignees were instructed to quarantine any affected product and return them to Cook Medical. Distributors were instructed to contact their customers and coordinate the return of the affected product to Cook Medical. Within 2 weeks of the initial recall notification, Level B effectiveness checks (40% of the total number of consignees) will be conducted. These custome
Quantitysee line one
DistributionWorldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, WA, WI, WV, and WY; and the countries of Algeria, Australia, Austria, Belgium, Canada, Chile, Cyprus, Den
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K944220WILSON COOK MULTIPLE BAND LIGATOR
Wilson-Cook Medical, Inc., Jan 24, 1995