Recall Z-0331-2011 ยท HSB Rod, Fixation, Intramedullary And Accessories

Stryker Howmedica Osteonics Corp.

Nail Holding Screw; 1 unit per package Stryker Trauma GmbH; Distributed by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 The nail holding Screw is used to assemble the nail to the target device. The Nail holding Screw is further the mating part (inner thread) of the External Compression Device as the instruments needs to be threaded into the Nail holding Screw in order to achiev

DateNov 9, 2010 (initiated Aug 30, 2010)
StatusTerminated
ReasonVariations in the outer diameter of the shaft of the External Compression Device and the inner diameter of the Nail Holding Screw may make assembly of these 2 instruments not possible.
Root causeOther
ActionStryker sent Recall notification letters by Federal Express on September 9, 2010 to all Stryker branches/agencies and Risk Managers, Chief of Orthopaedics of affected hospitals and Surgeons on September 10, 2010, that used the product. The letter identified the product, the problem, and the action that customers should take. Customers were instructed to: Examine their inventory and hospital locations to identify the affected product and return it to their branch or agency warehouse for reconciliation. Reconcile all product on the attached Product Recall Acknowledgment Form and fax a copy to
Quantity22
DistributionNationwide Distribution including AZ, CA, CO, FL, GA, IL, KY, MD, OH, OK, and TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.