Recall Z-0131-2011 ยท NKB Thoracolumbosacral Pedicle Screw System

Ebi, Llc

ARRAY and POLARIS Spinal System Medium Cross Connector Non Sterile Biomet Spine, Parsippany, NJ. 07054

DateOct 22, 2010 (initiated Sep 16, 2010)
StatusTerminated
ReasonOne of the two connector ends of the medium cross connector may not be fully secured to the rod after tightening. This may not be readily apparent to the operating surgeon, resulting in inadequate torsional stability.
Root causeOther
ActionBiomet Spine issued an Urgent Medical Device Recall letter dated September 16, 2010 to distributors and physicians, identifying the affected device and actions to be taken. Customers and distributors were requested to identify, remove, and immediately return the affected product to Biomet, following enclosed Product Return Instructions, and complete and return the Removal Tracking and Verification Form. Customers and distributors can contact Biomet at 1-800-526-2579.
Quantity82 units
DistributionWorldwide Distribution: USA and the countries of Australia, Columbia, Great Britain, The Netherlands, New Zealand, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062685EBI ARAY SPINAL SYSTEM
Ebi, L.P., Oct 6, 2006