Recall Z-0143-2011 ยท ITX Transducer, Ultrasonic, Diagnostic

Verathon, Inc.

The label (which is affixed to ScanPoint Communications Cradle) states: "...VERATHON ScanPoint Docking Station...Part No. 0570-0168...Bothell, WA USA..."

DateOct 26, 2010 (initiated Sep 20, 2010)
StatusTerminated
ReasonIncorrect programming causes QuickPrint to fail to recognize the communication Cradles. As a result, ScanPoint Communications Cradles could not communicate with ScanPoint Online Tool while using the QuickPrint interface. This will affect the transfer of exams and device calibration through QuickPrint interface.
Root causeSoftware design (manufacturing process)
ActionOn 09/20/10, Verathon issued an URGENT: Medical Device Accessory Recall Notification letter dated September 20, 2010 to consignees, identifying the affected product, the issue prompting the recall, and actions to be taken by consignees. The consignees are instructed to return any cradles within the serial number range between CC012000-CC012084, and the Part Number 0570-0168. The firm will provide a replacement communication cradle at no cost. Consignees can contact Verathon Customer Care at 425-867-1348 or 800-2331-2313 or send e-mail to cservice@verathon.com.
Quantity70 units
DistributionWorldwide Distribution: U.S. nationwide including states AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, KY, LA, MA, MD, MI, MN, MO, ND, NY, OH, PA, SC, TN, TX, UT, WI, and Washington DC, and the countries of Australia, Canada, France, South Korea, Taiwan, The Netherlands, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022153BLADDERSCAN BVI 6100
Diagnostic Ultrasound Corp., Aug 15, 2002
K030763DIAGNOSTIC ULTRASOUND BLADDERMASS BVM 6500 ULTRASOUND SYSTEM
Diagnostic Ultrasound Corp., Mar 26, 2003