Recall Z-1004-2011 ยท NQQ System, Imaging, Optical Coherence Tomography (Oct)

Optovue Inc.

Optovue iVue optical coherence tomography system, computer controlled opthalmic imaging system, manufactured by Optovue, Fremont, CA Computer controlled opthalmic imaging system. Device scans the patient's eye using low coherence interferometer to measure the reflectivity of the retinal tissue. The cross sectional retinal tissue structure is composed of a sequence of A-scans. It has a traditional

DateFeb 1, 2011 (initiated Jul 20, 2010)
StatusTerminated
ReasonUse of the CAM attachment may result in a positioning anomaly and mistreatment in the eyes of the patient.
Root causeSoftware design (manufacturing process)
ActionOptovue Inc. sent a Field Modification/Correction Notice dated July 20, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were informed that they could continue to use the iVue until the correction has been made. Customers were instructed to review any existing iVue patient data to ensure that this situation and an improper designation of eye assignment has not occurred. Upon completion of the updates, customers were to sign the attached form indicating they acknowledge that their iVue has been updated and
Quantity22 devices in the US.
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K091404IVUE, MODEL 100
Optovue, Inc., Apr 22, 2010