Recall Z-0144-2011 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

AXIOM Artis Systems VB23A-VB23H with Perivision functionality, Model Numbers: 5904433,5904649, 5904466, 5904441, 5904656, 7728350, and 7007755 Angiographic X-ray System

DateOct 26, 2010 (initiated Sep 9, 2010)
StatusTerminated
ReasonSporadic system crash while Perivision is performed.
Root causeSoftware design
ActionSiemens issued an Urgent Field Safety Notice to all customers. The letter informed customers of the potential issues and provided additional information on the implemented resolution. To resolve the problem, a system update was to be installed on each affected system under Update Instructions AX054/10/S. For questions phone (610) 448-3237.
Quantity286 units
DistributionUSA - Nationwide Distribution