Recall Z-0607-2011 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Beckman Coulter Inc.
Access Immunoassay Systems, Part Number: 81600 ; Access 2 Immunoassay Systems, Part Number: 81600N; Synchron LXi 725; UniCel DxI 800 Access Immunoassay Systems, Part Number: 973100; UniCel DxI 600 Access Immunoassay Systems, Part Number A30260; UniCel DxC 600i, Part Number: A25656; UniCel DxC 660i, Part Number: A64871; UniCel DxC 860i, Part Number: A64935; UniCel DxC 880i, Part Number: A59102; SY
| Date | Dec 13, 2010 (initiated Aug 13, 2010) |
| Status | Terminated |
| Reason | This recall was initiated because Beckman Coulter has received reports of erroneous test results being reported as a result of assay reagent packs being transferred between systems. Erroneous results will be generated as a result of missing or incorrect assay reagents. In these cases both false positive and false negative results are possible, depending upon the assay configuration. |
| Root cause | Pending |
| Action | Beckman Coulter, Inc. sent two (2) Product Corrective Action (PCA) letters (Access letter and DxI letter) with attached Customer Response form the week of August 09, 2010 to the affected customers. The letters were dated August 13, 2010, and identified the product, the problem, and the action to be taken by the customers. The letter reminded customers that it is critical that they load reagent packs according to the procedure provided in their Access or Access 2 Operators Guide, Instructions for Use, or Help system. A copy of the procedure was included with the letter. Customers were remind |
| Quantity | 9571 units (instruments) |
| Distribution | Nationwide Distribution & Canada |