ClearanceBrief
Recall Z-0570-2011 ยท KWT Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented

Zimmer Inc.

ZIMMER TRABECULAR METAL SHOULDER INSTRUMENTATION, DISTAL PILOT used for implantation of the Zimmer Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Shoulder Systems: Catalog Number 00-4310-008/018-13. The distal pilot manual surgical instrument is used in the surgical procedure to implant the Zimmer Trabecular Metal Humeral and Zimmer Trabecular Metal Reverse Shoulder Systems. The

DateDec 9, 2010 (initiated Sep 10, 2010)
StatusTerminated
ReasonThe results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant, delay in surgery, humeral fracture, or the need for cement in a press-fit application.
Root causeDevice Design
ActionZimmer, Inc. sent an URGENT DEVICE CORRECTION letter dated September 10, 2010 to surgeons, risk managers, and distributors. The letter identified the product, the problem, and the action to be taken. Sales Representatives were instructed to: a. provide the Dear Surgeon (Urgent Device Correction) letter and revised surgical technique directly to surgeons and risk managers at facilities using the device or b. provide a list of facilities and surgeons using the TM Humeral and TM Reverse shoulder systems to the Zimmer Recall Coordinator, and c. destroy previous versions of the surgical techniq
Quantity9,790 total distal pilots in distribution.
DistributionWorldwide Distribution - USA, Nationwide, Switzerland and Eschbach, Germany, Switzerland, Australia, China, India, Japan, Korea, Malaysia, Singapore, Taiwan, Brazil, Venezuela, El Salvador, Colombia, Canada, and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041549ZIMMER TRABECULAR METAL HUMERAL STEM
Zimmer, Inc., Sep 1, 2004
K050761ZIMMER TRABECULAR METAL HUMERAL STEM, SIZES 6MM AND 8MM
Zimmer, Inc., Apr 11, 2005
K052906ZIMMER TRABECULAR METAL REVERSE SHOULDER SYSTEM
Zimmer, Inc., Dec 19, 2005
K060704ZIMMER TRABECULAR METAL REVERSE SHOULDER SYSTEM, SIZES 8MM AND 10MM
Zimmer, Inc., May 19, 2006