Recall Z-0076-2011 ยท IYE Accelerator, Linear, Medical

Impac Medical Systems Inc

Sequencer Verify & Record System, for use with radiotherapy treatment machines, Class IIb, manufactured by Impac Medical Systems, Sunnyvale, CA.

DateOct 18, 2010 (initiated Sep 2, 2010)
StatusTerminated
ReasonSoftware issue-- matching criteria can be configured that permits an invalid patient match, resulting in overwriting of patient records with information from a different patient.
Root causeSoftware design
ActionElekta Impac Software issued a Important Safety Notice letter dated September 2010 to all customers identifying the software issue and clinical impact. Customer Support representatives will contact each site and modify site configuration or upgrade existing software to the version containing the corrective fix. Customers can contact Impac Software at 800 488-4672
Quantity797 units
DistributionWorldwide Distribution: throughout the US and the countries of Australia, Austria, Belgium, Canada, China, Hong Kong, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Japan, Jordan, Morocco, Netherlands, New Zealand, Poland, Portugal, Singapore, Sp
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K981313SEQUENCER
Impac Medical Systems, Inc., Jul 9, 1998