ClearanceBrief
Recall Z-0157-2011 ยท FPA Set, Administration, Intravascular

Baxter Healthcare Corp.

Baxter Non-DEHP I.V. Catheter T-Connector Extension Set, 6.5" (16.5 cm), 0.5 mL, PosiFlow Positive Displacement Access Device for I.V. Access, Male Luer Lock Adapter with Retractable Collar; an Rx sterile device; Baxter Healthcare Corporation, Deerfield, IL 60015 USA; product 2N9057 For the administration of drugs and solutions to a patient's vascular system through a vascular access device.

DateOct 27, 2010 (initiated Sep 9, 2010)
StatusTerminated
ReasonDuring packaging testing, Baxter observed tears and pinholes in Chevron-style pouches for certain infusion disposable products. The pinholes and tears could compromise the sterile barrier properties of the pouch.
Root causePackage design/selection
ActionBaxter Healthcare Corporation sent an Urgent Product Recall letter dated September 9, 2010, to all customers. The letter identified the product, the problem and the action to be taken by the customer. The customers were requested to examine their inventory in all locations to determine if they had any of the affected lots of the affected product codes and to remove the affected product from their inventory; they were to call the Baxter Healthcare Center for Service at 1-888-229-0001 to return the affected product, request a credit, or request replacement product; complete the attached custom
Quantity104,700 sets
DistributionWorldwide Distribution - U.S., Canada, Japan, Australia, New Zealand, and Chile.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K964918NIMA NEEDLELESS INJECTIONSITE MASTER ADAPTER WITH POSIFLOW POSITIVE DISPLACEMENT FEATURE, AND I.V. SETS
Critical Device Corp., Apr 14, 1997