ClearanceBrief
Recall Z-0022-2011 ยท KPR System, X-Ray, Stationary

Philips Medical Systems North America Co. Phillips

The BABIX holder is an accessory used with the DigitalDiagnost X-ray system. It is used to hold the baby either in a secured or immobilized standing position in front of the X-ray detector during the examination.

DateOct 7, 2010 (initiated Sep 9, 2010)
StatusTerminated
ReasonPotential of broken strands in the main steel cable supporting the accessory holder used with BuckyDiagnost and DigitalDiagnost X-ray systems.
Root causeEquipment maintenance
ActionPhilips issued URGENT-Device Correction Notice letters dated September 9, 2010 to their consignees identifying the affected products and action to be taken by customers. The letter states that Babix holder used with Philips BuckyDiagnost and DigitalDiagnost X-Ray System - VS Wall Stand Risk of steel cable fracture in the support holder. The consignees are informed that there is a potential of broken strands at the main steel cable within the holder which cannot be recognized by the user. The firm will issue the Field Change Order (FCO) and the Service Engineer will replace the Babix holder. Co
Quantity1 unit
DistributionNationwide Distribution in the states of CA, GA, and TN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982795PHILIPS BUCKY VISION
Philips Medical Systems, Inc., Nov 24, 1998