ClearanceBrief
Recall Z-0070-2011 ยท FPA Set, Administration, Intravascular

Baxter Healthcare Corp.

Baxter Interlink System Buretrol Solution Set, 83" (2.1 m), 150 mL Burette Drip Chamber Filter Valve, 2 Injection Sites, Male Luer Lock Adapter, Large Bore 4-Way Stopcock with Rotating Male Luer Lock Extension Set, 38" (97 cm), Vol. 5.8 mL, Injection Site, Male Luer Lock Adapter, 60 DPM; a sterile Rx IV fluid pathway device; Baxter Healthcare Corporation, Deerfield, IL 60015 USA; product code 3

DateOct 15, 2010 (initiated Sep 3, 2010)
StatusTerminated
ReasonThere are leaks in the Drip Chamber Check Valve Subassembly of the Solution Set. A leak in a sterile pathway could allow microbial entry, which could lead to a blood stream infection.
Root causeEmployee error
ActionBaxter sent an "Urgent Product Recall" letter dated September 2, 2010, via first class mail to the direct account, to the attention of the Director of Materials Management, with a copy sent to the Director of Nursing, on the same date, informing them that leaks were detected in the Drip Chamber Check Valve Subassembly used in the affected lot of Interlink System Buretrol Solution Sets. A leak in a sterile pathway could allow microbial entry, which could lead to a blood stream infection. The account was requested to examine their inventory in all locations to determine if they have any of the
Quantity948 units
DistributionNationwide distribution: Wisconsin only.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K984381BURETROL SOLUTION SETS
Baxter Healthcare Corp, Feb 19, 1999