Recall Z-2592-2010 ยท FRN Pump, Infusion

Stryker Instruments Div. of Stryker Corporation

Stryker PainPump2 400 mL PainPump with Luer Lock Tubing Set, sterile, REF 540-350, Stryker Instruments, Kalamazoo, MI. Delivers controlled amounts of medication and narcotics.

DateSep 28, 2010 (initiated Sep 2, 2010)
StatusTerminated
ReasonAll of the failures lead to the patient receiving less medication than intended. Certain lots of the pain pump have a high failure rate associated with a cracked port housing which leads to internal or external leaks. Some lots have a high failure rate associated with E2 errors.
Root causeNonconforming Material/Component
ActionThe firm sent our Urgent Medical Device Recall Notices dated 7/12/2010 to Risk Managers and Materials Managers. The recall notice listed specific lots of PainPump2 and PainPump2 BlockAid that were under recall. Consignees were told to check stock for the recalled product, quarantine the product, and contact Stryker to arrange return shipment of the recalled product. On 9/2/2010 the firm sent out a second Urgent Medical Device Recall Notice. This notice expanded the scope of the recall to include all products distributed within the past 12 months. The shelf life of the product is 1 year -
Quantity54,568 all products
DistributionWorldwide Distribution -- US, Canada, and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042405PAINPUMP1; PAINPUMP2
Stryker Corp., Oct 14, 2004
K043466PAINPUMP2
Stryker Instruments, May 10, 2005