ClearanceBrief
Recall Z-2467-2010 ยท ING Elevator, Wheelchair, Portable

The Braun Corporation

Braun wheelchair lift, model NL, Braun Corporation, Winamac, IN A wheelchair lift for motor vehicles.

DateSep 22, 2010 (initiated Aug 15, 2007)
StatusTerminated
ReasonThe wheelchair lifts threshold warning system may not operate and the wheelchair user can fall from the vehicle and be seriously injured.
Root causeDevice Design
ActionThe Braun Corporation issued a Service Bulletin 34273 entitled "Lift Threshold Sensitivity Bulletin 34273 on October 2007. Braun Corporation also sent a "Safety Recall" letter to the Braun lift owner and letters to all customers on or about August 15, 2007. The letters described the product, problem and the action to be taken by the customer. The customers were instructed to notify the National Highway Traffic Safety Administration (NHTSA) and conduct a safety recall if they used recalled lifts as original equipment in a system or vehicle that they built; ensure that any associated customers
Quantity8,356
DistributionWorldwide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K952946VANGATOR II
Crow River Industries, Inc., Nov 29, 1995