Recall Z-0623-2011 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Greiner Bio-One North America, Inc.

Vacuette 2 ml Lithium Heparin, Collection Tube, Manufactured by: Greiner-Bio One North America, Inc., 4238 Capital Drive, Monroe, NC. Venous blood collection tubes.

DateDec 16, 2010 (initiated Aug 25, 2010)
StatusTerminated
ReasonLithium heparin solution was contaminated by EDTA K (potassium) solution due to an operator error and a misinterpretation of a test result. Contaminated tubes will produce inaccurate test results.
Root causeOther
ActionGreiner Bio-One instructed customers to stop distribution and to complete an attached Product Disposition form and return it via fax to Greiner Bio-One. They were further instructed to destroy any product they may have. Distributors were instructed provide a list of their customers that have received the affected item via Excel file so that they can be notified of the recall.
Quantity18 cases (total 21,600 pieces)
DistributionNationwide Distribution: USA, including the states of CA, FL, IL, IN, KS, KY, ND, NE, NH, OH, SC, SD and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K960857GREINER VACUETTE BLOOD COLLECTION TUBE W/LITHIUM HEPARIN & GEL SEPARATOR
Greiner America, Inc., Aug 30, 1996