Recall Z-0624-2011 ยท FQO Table, Operating-Room, Ac-Powered

Steris Corporation

Cmax Surgical Tables, Model #'s: Cmax110, Cmax 220 manufactured by Steris Corporation in Montgomery, AL

DateDec 16, 2010 (initiated Aug 12, 2010)
StatusTerminated
ReasonCmax Surgical Tables may lose calibration if the table is operated only on battery power and if the table is in a low battery charge condition. This could result in the table moving into an unexpected tilt position when the Level button is depressed on the hand control keypad.
Root causeOther
ActionSteris issued Urgent Field Correction Notice letters dated August 26, 2010 to each affected consignee notifying them of the voluntary field correction. A Steris Representative will contact customers to arrange for STERIS to upgrade the software in the table. For further information about this corrective action, contact Steris at 1 800 333-8828.
Quantity1,635 tables
DistributionWorldwide Distribution: USA in the following states: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WV, and in the countries of Australia, Canada, China, Columb
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.