Recall z-2640-2010 ยท LFL Instrument, Ultrasonic Surgical

Stryker Instruments Div. of Stryker Corporation

Sonopet Ultrasonic Surgical System, Model UST-2001, Manufacturer Miwatec Co. LTD, Tokyo, Japan.

DateOct 1, 2010 (initiated Jul 15, 2010)
StatusTerminated
ReasonStryker Instruments acquired the Sonopet Ultrasonic Surgical System from Mutoh co. Ltd on 3/29/2010. Stryker determined that the Sonopet consode did not meet the IEC-60601-1 standard. Stryker decided to recall the Sonopet consode.
Root causeDevice Design
ActionThe firm to send URGENT MEDICAL DEVICE RECALL NOTIFICATIONS to customers dated 7/15/2010. The customers were instructed to continue to use the device. Stryker technicians were dispatched to replace a board with a new board.
Quantity1161
DistributionWordwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.