Recall Z-0024-2011 ยท KRL Detector, Bubble, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Air bubble detection cable assembly, catalog numbers 78-8067-7375-6 and 149892, Terumo Cardiovascular System, Ann Arbor, MI

DateOct 8, 2010 (initiated Oct 15, 2010)
StatusTerminated
ReasonMalfunction of the air bubble detection cable assembly causes a false air alarm or a check sensor message during system set-up. In some instances, users were not able to reset the alarm to allow for further air bubble detection. .
Root causeOther
ActionThe firm sent URGENT MEDICAL DEVICE CORRECTION notices dated September 17, 2010. The notices told consignees that a Terumo representative would visit, inspect the cables and replace the recalled cables. if the customer has a service contract, the visits would occur at the next scheduled preventative maintenance visit. Customers without a service contract will be contacted by Terumo representative to schedule an appointment. Customers can contact Terumo Customer Service at 800-521-2818.
Quantity5,000 estimated
DistributionWorldwide Distribution: USA states and territories of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, WA, WI, WV, and WY, and the countries of Australia, Belgium, Canad
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.