Recall Z-0132-2011 ยท DPW Flowmeter, Blood, Cardiovascular

Escalon Medical Corp

PD Access Vascular Access Device. Escalon Vascular Access, Inc., 2440 South 179th Street, New Berlin, WI 53146: Model 78050 18 Gauge GA Doppler Guided Bare Needle Assembly, 2 3/4 in Length, 0.038 in guidewire, Sterile EO, PD Access Vascular Access Device. Model 78060 Extended 18 Gauge Doppler Guided Bare Needle Assembly, 3 1/2 in Length, 0.038 in Guidewire, Sterile EO, PD Access Vascular Access

DateOct 22, 2010 (initiated May 5, 2010)
StatusTerminated
ReasonThe packaging of the vascular access device may potentially become damaged, compromising the sterility of the product. Use of a non-sterile product increases risk of patient infection. This recall involes: Model 78050 18 Gauge PD Access Vascular Access Device, Model 78060 Extended 18 Gauge PD Access Vascular Access Device, Model 75010 18 Gauge SmartNeedle Vascular Access Devices with Expiration Da
Root causePackage design/selection
ActionEscalon Vascular Access issued an "Urgent: Recall" letter dated May 5, 2010 to customers. The letter identified the affected product being recalled, described the problem, and the actions to be taken.. Consignees were recommended to immediately examine their inventory, quarantine the product, notify their customers, and return product affected. They also advised them that if they have further distributed the product, to please identify the customers and notify them at once of the product recall. Replacement product can be ordered from Vascular Solutions, Inc at 888-240-6001. Customer service
Quantity188,375
DistributionWorldwide Distribution: USA, including the states of AL, AR, AK, AZ, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL IN, IA, KS, KY, LA, ME, MD.MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, MN, NY, NC, ND, OH, OK OR, PA, RI, SC, SD,TN, TX UT, VA, WA, WV, and WI, and the countries of: Australia, Austria, Bahamas,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K903625DOPPLER NEEDLE AND FLOW MONITOR