Recall Z-2462-2010 ยท CGA Glucose Oxidase, Glucose

Abbott Point of Care Inc.

GCX Mountable Downloader-Recharger; (accessory to the i-STAT Portable Clinical Analyzer (model 300). The GCX Mountable Downloader-Recharger (GCX) is a connectivity component that is intended to be mounted to a GCX compatible mount or "arm" and provide a data connection to an authorized manufacturer's patient monitoring systems.

DateSep 23, 2010 (initiated Aug 20, 2010)
StatusTerminated
ReasonReliability issues associated with broken connection pins within the downloader that mate with the pins from the analyzer.
Root causeDevice Design
ActionAbbott Point of Care, Inc. sent Urgent Recall Notices, dated August 20, 2010, by Federal Express to two Clarion Health locations. The letter identified the product, the issue, and the actions to be taken by the firm and the customers. 1) An Abbott Point of Care representative will contact the customers to plan the replacement of affected product at their facility. 2) Replacement will occur in batches of approximately 50 units at a time. 3) Abbott will provide personnel to perform the replacements to help expedite the replacement. 4) The customer is required to acknowledge that all units
Quantity277
DistributionNationwide Distribution - USA, including the state of Indiana
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001387I-STAT PORTABLE CLINICAL ANALYZER, MODEL 300
I-Stat Corp., Oct 5, 2000