Recall Z-0018-2011 ยท CBK Ventilator, Continuous, Facility Use

Draeger Medical Systems, Inc.

Oxylog 3000, Emergency and Transport Ventilator; Catalog #2M86300; Drager Medical Inc., 3135 Quarry Road, Telford, PA 18969.

DateOct 7, 2010 (initiated Aug 19, 2010)
StatusTerminated
ReasonPatients received insufficient ventilation. The Oxylog 3000 Instructions are inadequate relating to dead space volume.
Root causeError in labeling
ActionThe firm sent Important Safety Notice Letter, dated August 2010, and Instructions for Use supplement to consignees. The letter explained the issue with the affected product. The letter asks customers to attach the Instructions for Use amendment sheet supplied with the letter to their corresponding Instructions for Use. The firm also states that a they will provide a dedicated pediatric breathing circuit for the Oxylog 3000. Questions regarding the letter should be directed to Michael Kelhart at 1-800-543-5047. Questions regarding the use of the device should be directed to DragerService T
Quantity78 potentially afftected units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062267OXYLOG 3000
Drager Medical B.V., Oct 31, 2006