ClearanceBrief
Recall Z-0009-2011 ยท BSZ Gas-Machine, Anesthesia

GE Healthcare, LLC

GE Datex-Ohmeda Avance Anesthesia System, manufactured by GE Healthcare, P.O. Box 900, FIN-00031 GE, Finland

DateOct 6, 2010 (initiated Aug 30, 2010)
StatusTerminated
ReasonUsers may inadvertently leave a test plug in the inspiratory flow sensor after completing Low P Leak Check on the SCGO (Switched Common Gas Outlet) configuration of Avance, Amingo, and Aisys anesthesia machines that may impact patient safety. If the test plug remains in place and therapy is initiated, the machine will not provide gas to the patient. This would inhibit mechanical or manual ventila
Root causeEmployee error
ActionGE Healthcare issued an " Urgent Medical Device Correction" letter dated August 30, 2010 to consignees. The letter was addressed to Chief of Anesthesia, Health Care Administrator/Risk Manager and Director of Biomedical/ Clinical Engineering. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Instructions. Customers were instructed to: 1. Ensure the plug is removed at the completion of the Low P Leak check. 2. Review the Preoperative Checkout procedure listed in section 4 of the User Reference Manual with appropriate personnel.
Quantity4541
DistributionWorldwide Distribution: USA: All States, including DC, except Delaware, and the countries of YEMEN, VIETNAM, VENEZUELA, UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, THAILAND, SYRIA, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SERBIA, SAUDI, ARABIA, RUSSIA, PORTUGAL POLA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081844GE DATEX-OHMEDA AVANCE
Datex-Ohmeda, Inc., Aug 29, 2008