Recall Z-2527-2010 ยท FMI Needle, Hypodermic, Single Lumen

Multi-Med, Inc.

All Med Medical 22ga x .75" Standard Bore Winged Administration Set Model # M2275S Reorder # 50-2223

DateSep 29, 2010 (initiated Aug 30, 2010)
StatusTerminated
ReasonHuber needles may core and result in port leakage or emboli being flushed our of the port.
Root causeDevice Design
ActionMulti-Med issued an Urgent: Medical Device Recall notification on 8/30/10 via fax and followed by hard copy letter. Users were advised of the problem and asked to immediately examine their inventory and quarantine the affected product. They were to also notify their customers of the recall. The firm stated in the letter that they would contact customers to determine the number of product that they had and the amount they expected from the field. The firm will coordinate a plan for the return of the affected product. Questions should be directed to Sue Starkey at 603-357-8733.
Quantity204 units
DistributionWorldwide Distribution -- USA and Mexico.
FDA recordOpen on fda.gov