ClearanceBrief
Recall Z-2343-2010 ยท FNJ Bed, Manual

Stryker Medical Div. of Stryker Corporation

Service parts for Crank Screw Assembly (PN1069-042-010S and 1069-042110S) Stryker Medical 3800 E. Centre Ave Portage MI 49002. Intended use: Service parts are replacement parts for the eye surgery stretcher.

DateSep 1, 2010 (initiated Feb 26, 2010)
StatusTerminated
ReasonThe fowler of the bed is raised by a crank. The crank consists of a steel screw that moves through a brass nut. Over time the screw can wear the threads on the brass nut. The screw and the nut can become misaligned, causing a loud squealing noise when the screw comes in contact with the drive tube. In addition the fowler can become inoperable over time with excessive wear of the brass nut. T
Root causeDevice Design
ActionThe firm, Stryker Medical, sent an "Urgent Medical Device Correction" letter dated February 26, 2010, to all customers. The letter describes the product, problem and action to be taken by customers. The letter states that Stryker field service tech will contact the firm and perform device corrections within four months. The customers were instructed to locate the recalled stretchers and verify that there is no grinding noise when activating the fowler. If the fowler exhibited a grinding noise, the customers were instructed to remove the stretcher from service and contact the recalling fi
Quantity226
DistributionWorldwide distribution: USA and countries including Australia, Canada, Chile, China, France, Germany, Greece, Spain, Italy, Japan, South Africa, Taiwan, Korea, Singapore, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.