Recall Z-2381-2010 ยท FRN Pump, Infusion

Hospira Inc

Hospira Inc. Symbiq One- and Two- Channel Infuser, List Numbers 16026 and 16027 respectively.

DateSep 8, 2010 (initiated Mar 26, 2010)
StatusTerminated
ReasonIf the administrator set is removed prior to the cassette carriage fully opening without closing the slide/roller clamp, unrestricted flow and delivery may result.
Root causeDevice Design
ActionHospitra Urgent Device Correction Notification mailed on 03/26/2010 via federal express. Tracking done by delivery tracking. Distribution of product continues and has not stopped. Additional notification was issued in 09/2010. A software correction is planned.
Quantity28485 units
DistributionWorldwide distribution: USA, Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Puerto Rico, Singapore and the UAE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041550HOSPIRA PHOENIX INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWARE