ClearanceBrief
Recall Z-0315-2011 ยท KWY Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented

Ortho Development Corporation

Distal Centralizer 10 mm

DateNov 3, 2010 (initiated Jul 28, 2010)
StatusTerminated
ReasonThe hip stem did not fit the distal centralizer because the hole was not drilled with the appropriate taper.
Root causeNonconforming Material/Component
ActionOrtho Development sent a Field Notification Memo dated July 28, 2010 to Japan Medical Dynamic Marketing instructing them to reconcile parts that may have been used and return unused parts to Ortho. Customers can contact Ortho Development at 1-801-619-3419.
Quantity39 units
DistributionWorldwide Distribution: Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K962448PRIMALOC CEMENTED HIP SYSTEM
Ortho Development Corp., Sep 19, 1996